Dry Air Generator can meet the gas requirements of various types of domestic and imported atomic absorption instruments, evaporative light scattering detectors, and multiple gas chromatographs, replacing high-pressure steel cylinders in the laboratory. It is especially suitable for gas supply in central control laboratories. It adopts a piston compressor with oil-free lubrication, low noise, low power, large flow, long life, and easy maintenance.
Precautions for using the equipment:
1. Remove the instrument from the packaging box and check for any damage caused by transportation. Verify that accessories and the certificate of conformity are complete.
2. If the compressor does not start during operation and the thermal protection relay activates, it indicates that the compressor temperature is too high. It will automatically return to normal after cooling.
3. The instrument should be installed on a vibration-free workbench. The surrounding air should be free of dust and corrosive gases. The ambient temperature should be 0-40°C, relative humidity ≤85%, and the power supply voltage 220V±10%, 50Hz.
4. To avoid excessive water accumulation inside the machine affecting air purity, press the drain switch for a few seconds before shutting down each time. (You can place a plastic bottle under the drain to prevent water splashing.) Drain at least once a day.
5. Observe the color-changing silica gel through the observation window. Users can replace the silica gel with new one or perform drying treatment as needed.
6. If the compressor does not start during operation and the thermal protection relay activates, it indicates that the compressor temperature is too high. It will automatically return to normal after cooling.
Note: The above information and technical parameters are compiled and released by Yangzhou Darui Electric Co., Ltd.